CRS AFS Purifier I
13
Troubleshooting
Support and Repair
If the AFS Purifier I is still in the warranty period, please contact your
dealer for support. If the warranty period has expired, please visit
www.chromres.com
for information about how to purchase a replacement
or regenerated cartridge.
Condition
Leak.
Indicator changes
quickly.
Possible Cause
O-rings between
cartridge and manifold
are not leak tight.
Leak in gas line.
Contaminated Gas.
Recommendation
Remove and check the
manifold O-rings for dirt
or scratches. Inspect the
sealing surfaces on the
cartridge and manifold
for scratches. Replace if
necessary.
Leak in gas line.
Verify that the source
meets specifications as
listed under “Limits” on
page 3.
If the gas source is sup-
plying other filters as well,
check the condition of
these filters. Otherwise
check specification and
procedures for verifying
gas quality.