XII. Troubleshooting guide
Symptom/
Cause
Action
error message
Instrument won't switch on
Flat battery.
Recharge or replace battery.
Pump not running
The pump is a PC
Reconfigure with PC
configurable option. software.
No confidence beep
Function disabled.
Reconfigure with PC software.
Gas reading when no gas
Zero drifted.
Restart instrument in clean
present
air.
Unstable/inaccurate gas
Sensor failure
Do not use; exit hazardous
reading
area immediately. Return
instrument for recalibration
or sensor replacement.
Autozero failed
Zeroing in
Switch off and restart in
contaminated
clean air.
atmosphere
Cannot autozero due to
Zeroing in
Switch off and restart in
alarm
contaminated
clean air
atmosphere
Calibration expired
The calibration due Send for calibration
date has passed
Flow fail clear blockage
Sample tube is
Clear blockage and press
blocked with water button to restart pump
or dirt or kinked
LCD too faint/dark
Contrast setting
Adjust using Portables
wrong
PC software.
Tetra
Troubleshooting guide
Fatal/Auto shut
Service
Configuration
User alert
Calibration