C) Commands
Table 1.2 lists the commands for the 2083-218# and briefly describes them.
After a command is sent the 2083 sends a return “>” indicating the command has been received and executed.
General Command Format - The general command format is {CND...}, where:
{ = start byte
C = 1 character, either C (command) or S (status)
N = 1 character command or status request
D = 1 character or more of data (depends on command)
} = stop byte
*
PLEASE NOTE: The two character {aa}(00-31) prefix, in the table below,
should be used ONLY when RS-485, (OPTION-Q), is selected.
Table 1.2 2083-218# Comm
mands
Com
mands
Set Input Frequency Band
{aaCBx}
where:
• x = 1 to select Band 1 (950-1550 MHz) .
• x = 2 to select Band 2 (1450-2050 MHz) .
Set Output Gain
{aaCGxxx}
where :
• xxx = +15 to +45 dB
Set Remote On
#
Just # sign
Set Remote Off
{aaCR0}
Option E:
Set 10 MHz Reference Mode
{aaCEx}
where :
• x = 0 to Select Internal Reference Mode
• x = 1 to Select External Reference Mode
• x = 2 to Select Auto Reference Mode