The identification label is affixed on the bottom of the equipment (Fig. 1, page 6).
ATTENTION
- Removing the identification label and/or any other stickers from the product will cause automatic loss of
warranty.
HOW TO IDENTIFY YOUR ULTRASONIC CLEANER
Some procedures are necessary for the best functioning and durability of your equipment. The preventive
manintenance corresponds to the fulfilment of all procedures listed below.
Make sure the equipment is unplugged before performing any kind of maintenance.
PREVENTIVE MAINTENANCE
PERIODICITY
MAINTENANCE
DAILY
External cleaning:
Unplug the equipment. Clean it with a soft cloth moistened in water and neutral biodegradable
detergent, next, use another soft cloth moistened with alcohol 70% or peracetic acid at 1%.
Clean the lid the same way. Wipe it with smooth movements in only one way.
Vat cleaning:
Clean the stainless steel vat and change the cleaning solution after each cycle to maintain the
cleaning efficiency
. The contaminated solution must be disposed according to the instructions
of the topic “
How to Use the Cristófoli Ultrasonic Cleaner
”, instruction 10 (Page 10).
After disposing the water/solution, remove the excess of residues that may have accumulated
at the bottom of the vat with a soft sponge, cloth or even paper towels, then clean it with a soft
sponge with water and neutral detergent, rinse it with running water (be careful not to get the
control panel wet, avoiding this way that some water get into the equipment through any of its
many orifices), dry the vat with a soft cloth or paper towel and finish by disinfecting it with alcohol
70% three times according to the technique recommended for its use (BRASIL 1994).
ANNUAL
We recommend performing a general preventive maintenance of the equipment with a
Cristófoli authorized technician. See “
Authorized Technical Assistance
” (Page 13).
Table 3
11
CRISTÓFOLI
U L T R A S O N I C C L E A N E R
I N S T R U C T I O N M A N U A L