Cristofoli Mini-incubator Manual Download Page 3

03

The mini-incubator is not connected to the 

electric outlet;

Plug  the  mini-incubator  into  the  electric 

outlet;

See  “

How  to  Proceed  when  Service  is 

Needed

 (Page 4);

Printed fuse burned out;

No power;

Check  if  there  is  a  power  outage  in  your 

building;

SOLUTIONS

POSSIBLE CAUSES

THE MINI-INCUBATOR DOES NOT TURN ON

Defective circuit board;

See  “

How  to  Proceed  when  Service  is 

Needed

 (Page 4);

SOLUTIONS

POSSIBLE CAUSES

A MINI-INCUBADORA LIGA, MAS NÃO AQUECE

SOLUTIONS

POSSIBLE CAUSES

If the problem persists after the verification of all the items listed,

contact your local dealer.

The mini-incubator lid is open;

In case the lid of the incubation area is open, 

close it properly;

Defective circuit board;

See  “

How  to  Proceed  when  Service  Is 

Needed

 (Page 4);

A TEMPERATURA DA MINI-INCUBADORA SOBE EXCESSIVAMENTE

  All  Cristófoli  Mini-incubators  are  tested  and  monitored  individually,  according  to  the 
parameters of the table below.

QUALITY CONTROL

NOTE:

 Heating Time values are expressed considering the technical data chart regarding the 

environment  conditions,  (temperature  /  altitude).  Incubation  time  will  depend  on  the 

biological indicator used. (check the time needed for your BI by consulting its manufacturer 

instructions.

A MINI-INCUBADORA DEMORA PARA ATINGIR TEMPERATURA OU NÃO A MANTÉM

SOLUTIONS

POSSIBLE CAUSES

  The metallic identification label is located underneath the equipment. Its purpose is to 
identify the 

 technical data.

mini-incubator

ATTENTION!

 - Removing the Identification Label and/or any other stickers from the product will 

cause automatic loss of the warranty.

HOW TO IDENTIFY YOUR MINI-INCUBATOR

Fig.5

*  In  case  the  altitude  and/or  temperature  of  your  work  place  is  different  from  the  one 
mentioned in this manual, contact Cristófoli by the e-mail: [email protected].

Note:

 The manufacturer reserves the right to make changes and/or improvements to this 

product at any moment without prior notice.

CERTIFICATIONS ..........

CAPACITY .....................................................

WEIGHT.........................................................

CYCLE .........................................................

OVERALL CLEARANCE ..................................

CLEARANCE REQUIRED TO OPEN THE LID ......

EXTERNAL DIMENSIONS (H x D) .....................

LID DIMENSIONS (H x D) ................................

VOLTAGE .....................................................

FREQUENCY .................................................

POWER ........................................................

POWER CONSUMPTION ................................

MAXIMUM TEMPERATURE ..............................

*PROPER WORKING TEMPERATURE ................

*PROPER WORKING ALTITUDE .......................

4 biological indicators

240 g

24 o 48 hours (steam)

15 cm for each side

12 cm

8,7 x 7,6 cm

3,3 x 7 cm

90 - 253V AC single-phase

50/60 Hz

10 Watts

10 Watts hour

60 ºC

15 ºC to 40 ºC

Up to 2.000 m

TECHNICAL DATA CHART

CRISTÓFOLI MINI-INCUBATOR

Cristófoli Mini-incubators are 

manufactured by Cristófoli Biossegurança, 

company  which  Quality  Management  System  is  certified  by  ISO 
13485:2003, and in accordance with 

BPF - Boas Práticas de Fabricação - 

ANVISA (Brazilian standard similar to 

GMP - FDA / US).

TROUBLESHOOTING

 

WARNING! 

For any parts replacement, contact your local dealer or the manufacturer. We 

listed below some possible problems and solutions you can try on your own.

TECHNICAL DATA

24 or 48 hours

(according to the B. I.)

55 ºC to 60 ºC

Cold or hot Mini-incubator

60 minutes

PREVENTIVE MAINTENANCE

 

 

Some preventive procedures are necessary for the best functioning and durability of your

mini-incubator:

  Use 

only

 self-contained biological indicators, proper for steam sterilization;

 

incubator

the incubation area, use a cotton swab 

Keep the mini-

 clean. To clean 

 

with 

alcohol 70% or 

peracetic acid at 1% and rub inside the indicators compartments some 

times

 cause d

s.

. The use of other materials and/or products may

amage

 

The external cleaning must be done daily using a soft cloth and biodegradable neutral 

detergent, next, clean it thoroughly with a soft cloth and alcohol 70%or 

peracetic acid at 

1%.

Insufficient exposure time to the proper temperature;
Incorrect handling of the incubator or the biological indicators;
Inappropriate indicators for incubation (not proper for steam, incorrectly stored or overdue 
shelf life);
Removal of the lid, causing the incubation temperature to drop beyond the ideal level;
Lack of preventive maintenance;
Power outage;
Defective incubator.

POSSIBLE INCUBATION FAILURES

1

2

3

4

6

7

5

ATTENTION!

 - Failures during the incubation process are detected during monitoring.

must be repeated with a new control indicator.

AUTOCLAVE STERILIZATION MONITORING

  The purpose of the Biological indicators is to monitor autoclave sterilization cycles, they are 
considered the most accepted method for monitoring the sterilization process because they 
assess  it  directly  by  killing  known  highly  resistant  microorganisms  (e.g.,

  Geobacillus 

stearothermophilus

 for steam and 

Bacillus subtillis

 for ethylene oxide), rather than merely 

testing the physical and chemical conditions necessary for sterilization. Because spores used 
in  BIs  are  more  resistant  and  present  in  greater  numbers  than  the  common  microbial 
contaminants found on patient-care equipment, an inactivated

*

 BI indicates other potential 

pathogens in the load have been killed. The articles must have been previously cleaned 
before being sterilized, since the organic matter protects the microorganisms from the action 
of the sterilizing agents.
 

“Correct functioning of sterilization cycles should be verified for each sterilizer by the 

periodic  use  (at  least  weekly)  of  Bis”.  (Centers  for  Disease  Control  and  Prevention. 
Guidelines for Infection Control in Dental Health-Care Settings - 2003. MMWR;52 (No. RR-
17):  page  24).  Each  institution  must  also,  according  to  the  Brazilian  Health  Ministry, 
establish a monitoring routine of its autoclaves, with the objective of detecting faults in 
their equipment and /or process.

 

In  hospitals,  monitoring  must  be  done  on  a  daily  basis  (AORN,  1994).  A  biological 

indicator must be used for every sterilizer load that contains an implantable device. The 
load can only be released after the results of incubation are known.

 

The  number  of  test  indicators  to  be  put  in  each  piece  of  equipment  must  be 

established according to its volume. A control BI, from the same lot as the test indicator 
and not processed through the sterilizer, should be incubated with the test BI; the control BI 
should yield positive results for bacterial growth and the results from the test BI should be 
negative.  The  incubation  process  must  begin  from  15  minutes  after  the  end  of  the 
sterilization, and no later than two hours after it.

inactivated = negative).

Heating Time

Incubation Time

Temperature

CRISTÓFOLI

EQUIPAMENTOS DE BIOSSEGURANÇA LTDA

ROD BR 158 Nº 127 - CAMPO MOURÃO - PR - BRASIL

CNPJ 01.177.248/0001-95

INSCR. EST. 90.104.860-65

CEP 87303-327

TEL: 55 44 3518-3434

MADE IN BRAZIL

CAPACITY

4 Indicators

Cristófoli Mini-incubator

MODEL

50/60 Hz

S N

90 - 253V~ - 0,07 kW - 0,1A

VOLTAGE

RODUCT: INCUBATOR FOR BIOLOGICAL INDICATORS

USE INSTRUCTIONS, PRECAUTIONS, CONSERVATION AND STORAGE:

SEE INSTRUCTION MANUAL.

RESPONSIBLE TECHNICIAN

Eduardo Luiz Soppa

CREA-PR: 109201/D

ANVISA REGISTER

EXEMPT

According to Resolution RDC nº 260/02

LOT

FREQUENCY

Reviews: