![Creative PC-68B User Manual Download Page 21](http://html1.mh-extra.com/html/creative/pc-68b/pc-68b_user-manual_2680761021.webp)
User Manual for Wrist Oximeter
-16-
3.2.1 Record Interval Settings
On menu screen, select “Rec Interval” and press “
■
” key
to enter Record Interval Setting screen, as shown in Figure
13.
Figure 13 Record Interval Setting
Operation Instructions:
Short time press “
■
” key:
active the option;
Short time press “
▲
” key:
select record interval;
three options: “1s”, “4s” and “8s;
Short time press “
■
” key:
confirm the selection;
Longtime press “
■
” key:
back to the menu screen;
Note:
The length of data record is limited to 30 seconds
at least, and the maximal length for one record is also
limited to 60 hours (for 1 second interval) , 240 hours
(for 4 second interval) or 480 hours (for 8 second
interval).