background image

User Manual for Handheld Pulse Oximeter 

  29 

 

Summary of Contents for PC 66B

Page 1: ...tten consent Creative reserves the right to revise and amend it at any time without prior notice Version of the Manual Ver 2 1 Revised date October 18 2012 All rights reserved Notes The contents conta...

Page 2: ...ximeter in environment with inflammable gas such as some ignitable anesthetic agents DO NOT use the oximeter while the testee is under MRI or CT scanning Warnings An uncomfortable or painful feeling m...

Page 3: ...perating it When it is carried from a cold environment to a warm and humid environment please do not use it immediately DO NOT operate the button on the front panel with sharp materials High temperatu...

Page 4: ...llation of Battery and Holder 4 3 SpO2 Probe Connection 5 4 Operation 7 4 1 Default Screen 7 4 2 Display Screen with PI Value 9 4 3 Other Screen Layouts 9 4 4 Indication for No Signal 10 4 5 Menu Setu...

Page 5: ...es 2 Navigation key Up Left Increase Press this key the default screen can be shifted to display pulse rate PR or perfusion index PI When on the system setup screen press it to move the cursor upwards...

Page 6: ...parameter setting press this key to confirm 4 Data interface used for uploading data Optional function 5 Reserved port for future use 6 Power Back Power on off the device by longtime pressing short t...

Page 7: ...se Rate value Data transmission to PC for view and analysis Optional Audible Visual alarm function Power management function 1 5 Intended Use This Handheld Pulse Oximeter is intended for measuring and...

Page 8: ...Battery and Holder 1 Open the rear panel with coin or an ordinary flat screwdriver as shown in Figure 2 Figure 2 2 According to the polarity mark insert three AA batteries into battery house as shown...

Page 9: ...nnect the SpO2 probe to the right panel s connector labeled SpO2 After starting the monitor insert one finger into the probe index finger middle finger or ring finger with proper nail length according...

Page 10: ...uff for blood pressure measurement or during venous infusion 5 Do not let anything block the emitting light from the sensor 6 Vigorous exercise and electrosurgical device interference may affect the m...

Page 11: ...Description SpO2 99 SpO2 value 99 is a percentage value 10 12 45 Current time PR 65 Pulse rate value 65bpm Battery indication Data memory icon when the data on screen becomes stable the device will st...

Page 12: ...n Alarm Icon Pulse strength bar graph Lower part displays plethysmogram Note If the leadwire falls off during the measurement process screen will appear No Signal Oximeter gives off the sound alarm li...

Page 13: ...creen with PI value is shown below Figure 7 Display Screen with PI Value 4 3 Other Screen Layouts On default screen press Mode OK key to enter into other screen layouts as shown in Figure 8A Operation...

Page 14: ...creen 4 4 Indication for No Signal If No Signal prompts on screen as shown in Figure 9 please check whether the probe connects well and whether the finger is in appropriate position If the state of No...

Page 15: ...Description SETUP set parameter values refer to Chapter 4 5 1 for details UPLOAD DATA enter into uploading status refer to Chapter 4 5 2 for details VERSION for viewing version number of the software...

Page 16: ...User Manual for Handheld Pulse Oximeter 12 Figure 11 System Setup Screen A Figure 11 System Setup Screen B Figure 11 System Setup Screen C...

Page 17: ...lue and year value are the same with Month adjustment Date Format MM dd yy Note The setting operations of other parameters such as TIME CONTRAST ALARM SOUND etc are the same as date setting 2 TIME Tim...

Page 18: ...so limited to one hour for one second interval 2hours for 2 seconds interval 4 hours for 4 seconds interval or 8 hours for 8 seconds interval 4 5 2 Upload Data On the menu screen select UPLOAD DATA an...

Page 19: ...On the menu screen select VERSION and then press key for entering into version screen as shown in Figure 13 Figure 13 4 6 Data Recall On the default display screen longtime press Recall backlight key...

Page 20: ...in Figure 15A Figure 15A Trend Graph Recall Screen Screen description SpO2 The left ordinate value is SpO2 value 12 50 35 Measuring time 2s Record interval is 2 seconds PR The right ordinate value is...

Page 21: ...Graph Recall Screen Press key to view trend graph Press key to return to recall list screen 4 6 2 Data Deletion Choose one record in the recall list then longtime press key an icon will be displayed...

Page 22: ...hen finishing the selection of the records needed to be deleted press key then the screen will prompt that Delete all the marked records as shown in Figure 17 at this time select Yes and press OK key...

Page 23: ...Specifications Transducer dual wavelength LED sensor Measurement wavelength Red light 660 nm Infrared light 905 nm Maximal optical output power less than 2mW maximum average Measuring range 35 100 Me...

Page 24: ...erfusion condition The accuracy of SpO2 and PR measurement still meet the precision described above when the modulation amplitude is as low as 0 5 I Resistance to interference of surrounding light The...

Page 25: ...ingress of liquids Ordinary equipment without protection against ingress of water Electro Magnetic Compatibility Group I Class B 6 Accessories 1 A probe 2 A holder 3 Battery AA 3 4 A User Manual 5 A...

Page 26: ...pressure 50kPa 107 4kPa The oximeter is calibrated in the factory before sale there is no need to calibrate it during its life cycle However if it is necessary to verify its accuracy routinely the us...

Page 27: ...ctly inside and try again 2 Let the patient keep calm Can not turn on the device 1 The batteries are drained or almost drained 2 The batteries are not inserted properly 3 The device is malfunctioning...

Page 28: ...Pulse Oximeter 24 Appendix Key of Symbols Symbol Description Symbols on the screen SpO2 The pulse oxygen saturation PI Perfusion Index Pulse rate Unit beats per minute Low battery voltage Alarm Icon...

Page 29: ...r Power Back Key Mode OK Key Recall Backlight Key Navigation Key Data Interface CE mark Serial number Date of manufacture Authorised representative in the European community Manufacturer including add...

Page 30: ...e of Measurement Based on Lamber Beer law the light absorbance of a given substance is directly proportional with its density or concentration When the light with certain wavelength emits on human tis...

Page 31: ...e status of SpO2 can be determined in time by measurement and find the hypoxemia patient earlier thereby preventing or reducing accidental death caused by hypoxia effectively 3 Normal SpO2 Range and D...

Page 32: ...e is an arterial occlusion proximal to the sensor Blood vessel contraction caused by peripheral vessel hyperkinesias or body temperature decreasing 5 Factors causing low SpO2 Measuring value pathology...

Page 33: ...User Manual for Handheld Pulse Oximeter 29...

Reviews: