5. MAINTENANCE
You are advised to send us your instrument for maintenance (at CPF or an authorized
service centre) once a year.
Normal maintenance consists of checking the calibration and cleaning the pipes.
FOR CORRECT MAINTENANCE:
Always use original sensors or spare parts and do not have the instrument maintained at
unauthorized centres, otherwise the warranty shall be forfeited.
REPLACING THE BATTERY
The replaceable battery (BA045) is easy to replace.
Turn off the instrument and
disconnect it from the battery charger
then open the cover at the back of the
case
,
remove the battery and take off the connector.
Next, attach the polarized connector on the new battery, put the battery in the
compartment and re-close the cover.
6. TECHNICAL CHARACTERISTICS
Pressure range
0... +35bar
Resolution
0.1 bar
Accuracy
±1% fs
Display
LCD 8 characters on 2 lines
Battery
Li-ion 3.6Vdc 1.1 Ah
Average battery duration
30 hours
Battery charging time
4 hours
Pneumatic connection
M1 0x1
Weight of the case
Approx. 1.4 Kg
Dimensions of the instrument
260mm x 80mm x 35mm (excl. the sensor l =21 0mm)
Printer
External infrared printer; paper width 58mm
Power
230Vca external power supply for charging or direct
power (also for the printer).
Operating temperature
-5 +60°C
Storage temperature
-20 +85°C
Max applicable pressure
65 bar
AC999 conforms to the provisions of standard EN 3-7/2005