10-4
Valleylab FT10 Energy Platform User’s Guide
VLFT10GEN Specifications
Power Cord Specifications
This system is factory equipped with a 110 VAC hospital grade NEMA 5-15 power cord.
Should the AC power cord need to be replaced to match another plug configuration, the
replacement plug/cable/receptacle configuration must meet or exceed the following
specifications:
•
100-120 VAC
Cable - SJT16/3, IEC color code, maximum length 15 ft. (5 m)
Plug - minimum 10 A - 125 VAC
Unit receptacle - IEC female, minimum 10 A - 125 VAC
•
220-240 VAC
Cable - H05VVF3G1.0 VDE, maximum length 15 ft. (5 m)
Plug - minimum 6 A - 250 VAC
Unit receptacle - IEC female, minimum 6 A - 250 VAC
Backup Power
The VLFT10GEN retains all user programmed features, calibration, and statistical data
when switched off and unplugged. The VLFT10GEN operates within specification when
switched over to a supplied-line power by hospital backup systems.
Equipotential Ground Connection
An equipotential ground connection is provided to allow connection of the VLFT10GEN to
ground.
EKG Blanking and Smoke Evacuation
Interlink cable receptacles are provided to signal other devices, such as an EKG or smoke
evacuator, that the VLFT10GEN is active. The receptacle is a 2.5 mm mono jack. It is
electrically isolated from the internal ground referenced electronics with the shell
electrically connected to the chassis for ESD protection.
Important
Contact your local Covidien representative for alternative internationally approved power-cord
options.
Real time clock battery
Battery type – Lithium CR1620 or CR1632
Battery life – 75 mAh minimum