SurgiStat Low-Power Electrosurgical Generator Service Manual
6-3
Fuse Replacement
Fuse Replacement
Fuses for the unit reside directly below the power cable receptacle on the rear of the unit.
To replace the fuses, follow this procedure:
1. Unplug the power cord from the wall outlet.
2. Remove the power cord from the power cable receptacle on the rear panel.
3. To release the fuse drawer, insert a small flathead screwdriver into the slot on the
drawer below the power cord receptacle and slide the drawer out.
①
Fuse Drawer
4. Remove the two fuses (T3.15AL250V) and replace them with new fuses with the same
values.
5. Insert the fuse drawer into the power cable receptacle.
Use the following fuses:
SurgiStat
100-240 V
VAC
250
Amps
3.15 A
Type
Slow Blow
Size
5 x 20 mm
Important
If the unit does not display an error and does not power on, check fuses.
1