2.2.1 Gas Inlet System Overview
The gas inlet system, shown in Figure 2-3 and Figure 2-4, entrains room air. It also delivers
externally supplied oxygen to the ventilator and regulates the oxygen to a pressure usable by
the ventilator.
Figure 2-3. Gas inlet system pneumatic diagram
7-00243
Air intake
check valve
Pressure relief
check valve
Air intake filter
From
safety valve
To/from
piston/cylinder
assembly