General information
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
1-15
1.8 Tools, equipment, and service materials
The tools, equipment, and service materials listed in Table 1-7 are used to service the 700
Series
Ventilator.
Caution
Always use metric tools to remove metric fasteners. Using nonmetric tools to remove
metric fasteners can damage fasteners.
Table 1-7: Tools, equipment, and service materials
Description
Manufacturer and model or
Covidien part number
Where used
Hex drivers, 1.5-mm, 2.5-mm, 3-mm,
4-mm, and 5-mm
Local supplier
Various service procedures
Hex driver, 1/16-in.
Local supplier
Knob set screw
Hex driver, 7/64-in.
Local supplier
Autozero solenoid retaining screws
Nutdrivers or wrenches, 5-mm, 5.5-mm
(or 7/32-in.), 7-mm, 8-mm, and 11-mm
Local supplier
Various service procedures
Wrench, open-ended (or adjustable)
Local supplier
•
10-mm
•
Inspiration manifold thermistor
(previous version), nurse call
connector lock nut
(Communications option
assembly)
•
14-mm
•
Nebulizer connector lock nut
(Communications option
assembly)
•
19-mm
•
Inspiration manifold thermistor
(current version), oxygen
regulator pressure transducer;
oxygen regulator nut, remote
alarm connector lock nut
(Communications option
assembly)
•
3/16-in.
•
Oxygen regulator adjustment
screw
•
Serial (RS-232) port hardware
(Communications option
assembly)
•
3/8-in.
•
Oxygen regulator nut
POZIDRIV™*
screwdrivers, no. 0, 1, and 2
Local supplier
Various service procedures
NOTE:
Use POZIDRIV™* screwdrivers only to remove POZIDRIV™* screws. Using Phillips-head screwdrivers on
POZIDRIV™* heads may damage the screw heads.