Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-69
8.14.6 External battery harness
This section details correct installation of the revised external battery harness
(G-060113).When mounting the external battery connector into the rear panel of the
ventilator, verify the orientation. See Figure 8-56. When installing the harness, verify the
cables of the harness are connected per Table 8-1.
Figure 8-56. External battery connector
(See Figure 8-57). Remove the external battery harness as follows. Install by reversing removal
procedure.
1.
Remove power assembly (Section 8.14.2).
2.
Open connector cover.
3.
Disconnect external battery connector harness from BBU PCB.
4.
Remove two M3 x 8 flat-head, POZIDRIV™* screws that attach harness to power
assembly. Pull out harness through cutout in power assembly.
NOTE:
When installing the external battery harness, be sure to attach the ground wires to the
chassis with the harness retaining screw.
Table 8-1: External battery connections
Terminal (J2)
Cable
Signal name
1+
Large red cable
Bve
2+
Large black cable
Battery -ve
1-
Small black cable (with capacitor)
Extbat JM1
2-
Small black cable (without capacitor)
Extbat JM2
1+
1-
2+
2-
7-00429A
The harness ground
lug is mounted under
M3 barb nut
Internal view in the external
battery connector J2
Note orientation
and logo