General information
1-4
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
1.4 Configuration information
The 700 Series Ventilator System is available in a variety of versions, intended to meet
differing needs and regulations throughout the world. The major differences in
configuration among ventilators are listed below:
•
Language.
Keyboard (user interface) and other ventilator labeling available in assorted
languages.
•
Electrical requirements.
Available in 110 V and 230 V.
•
Power cord.
A variety of plug ends available.
•
Oxygen fitting and hose.
Available in DISS (diameter index safety standard) male and
female, NIST (non-interchangeable screw thread), Australian type, Air Liquide™*, and
Dräger™*.
•
Mounting.
Cart or shelf mount available.
•
Accessories.
Ventilators may have the accessories listed in Section 1.5.
1.5 Accessories
The following accessories are either required or can be used with the ventilator.
Ventilator breathing circuit.
A variety of reusable Puritan Bennett™ circuits, adult and
pediatric, with and without water traps, and with and without heated wire, is available.
Contact your Covidien representative. In addition, other breathing circuits may be used with
the ventilator, provided they ensure that the ventilator breathing system meets the
specifications stated in Table 1-1 (see
Results of ventilator breathing system testing
).
Humidification device.
The 700 Series Ventilator supports use of an optional humidification
device, including a heated humidifier, heat and moisture exchanger (HME), or heated wire.
A Fisher & Paykel™* MR730 Humidifier is available for the ventilator. Mounting brackets are
available for the Fisher & Paykel™* humidifiers and the Hudson RCI™* ConchaTherm™*
Humidifier.
1.6 Specifications
Ventilator specifications are listed in Table 1-1.
Table 1-1: Specifications
Physical characteristics
Weight
Ventilator only: 30 kg (66 lb)
Cart only: 18 kg (40 lb)
External battery: 12.75 kg (28 lb)
Dimensions
Ventilator only: 378 mm high x 515 mm wide x 370 mm deep
(14.9 in. high x 20.3 in. wide x in. 14.6 in. deep)
Ventilator and cart: 1255 mm high x 515 mm wide x 370 mm deep
(49.4 in. high x 20.3 in. wide x 14.6 in. deep)
Environmental requirements
Temperature
Operating: 5 to 45 °C (41 to 113 °F) at 10 to 95% relative humidity
Storage: -40 to 60 °C (-40 to 140 °F) at 10 to 95% relative humidity
Atmospheric pressure
Operating: 8.7 to 16.0 psi (600 to 1100 hPa)
Storage: 7.3 to 16.0 psi (500 to 1100 hPa)
Altitude
Operating: Up to 4570 m (15,000 ft)
Storage: Up to 15,240 m (50,000 ft)
Storage
Maintain under conditions listed above. Remove batteries from ventilator before storage.
See “Electrical specifications, Battery shelf life when charge is not maintained” for battery