Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-13
Figure 8-9. Removing shelf mount from ventilator
8.4.2 Removing/installing collector vial support bracket
(See Figure 8-8). Using 3-mm hex driver, remove two M4 x 10 screws, flat washers, and split-
ring washers that retain support bracket to shelf mount plate. Remove support bracket.
Install by reversing removal procedure.
8.4.3 Removing/installing humidifier support bracket
(See Figure 8-8). Remove two M5 x 10 flat-head screws (Fisher & Paykel™* bracket) or four
M5 x 20 flat-head screws (Hudson RCI™* bracket) that retain support bracket to shelf mount
plate. Remove bracket. Install by reversing removal procedure.
To unlock
Shelf
Ventilator
Shelf
Ventilator
7-00184
Interlocking
rails locked
in place
Interlocking
rails clear
slots on shelf