Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-9
8.3.4 Removing/installing cylinder covers
Using 7-mm nutdriver, remove four nuts with captive lockwashers that retain each cylinder
cover. Install by reversing removal procedure.
8.3.5 Removing/installing collector vial support bracket
(See Figure 8-4). Using 3-mm hex driver, remove two M4 x 8 screws, flat washers, and split-
ring washers that retain bracket to cart. Remove bracket. Install by reversing removal
procedure.
Figure 8-4. Removing collector vial and Fisher & Paykel™*
Humidifier support brackets from cart
8.3.6 Removing/installing humidifier support bracket
8.3.6.1 Fisher & Paykel™* humidifier support bracket
(See Figure 8-4). Using 4-mm hex driver, remove two M5 screws, flat washers, and split-ring
washers that retain bracket to cart. Remove bracket. Install by reversing removal procedure.
8.3.6.2 Hudson RCI™* humidifier support bracket assembly
(See Figure 8-5). Remove two M6 x 20 screws, split-ring washers, and flat washers that retain
bracket assembly to cart. Remove bracket. Disassemble bracket assembly as required. Install
by reversing removal procedure.
8.3.7 Removing/installing basket
(See Figure 8-6). Partially remove screw retaining one of two basket-retaining ball brackets.
Slip basket off studs on ball brackets and studs on inside of column. Install by reversing
removal procedure.
7-00179
M5 x 12 screw (2 PL)
Flat washer (2 PL)
Split-ring washer (2 PL)
Humidifier
support bracket
Collector vial
support bracket
M4 x 8 screw (2 PL)
Flat washer (2 PL)
Split-ring washer (2 PL)
Summary of Contents for Puritan Bennett 700 Series
Page 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...