Puritan Bennett™ 700 Series Ventilator System Service Manual
8.3.3 Removing/installing oxygen cylinder brackets
Remove the oxygen cylinder brackets from the cart as follows. Install by reversing removal
procedure.
1.
Loosen cylinder-retaining wing screws. Lift out cylinders.
2.
Using 7-mm nutdriver, remove four M4 nuts with captive lockwashers that retain each of
two lower cylinder supports. Remove supports.
3.
Using 5-mm hex driver, remove two M6 x 16 screws, flat washers, and split-ring washers
that retain each of two upper cylinder brackets. Remove brackets.
Figure 8-3. Removing oxygen cylinder brackets
7-00178
Oxygen cylinder
support bracket
(upper) (2 PL)
M6 x 16 screw (4 PL)
Flat washer (4 PL)
External lockwasher
(4 PL)
Nut with captive
lockwasher (4 PL)
Wing screw
(4 PL)