Performance verification
5-4
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
Clean and inspect the ventilator as follows.
1.
Clean ventilator exterior.
2.
Remove any water from the humidifier jar and dry, if applicable.
3.
Open and prop UI.
4.
Using ESD-safe vacuum cleaner, clean interior of ventilator.
5.
Remove and inspect main fan filter. Clean, vacuum, or replace as required.
6.
Inspect outer gasket around lid to ensure it is not torn and that entire gasket is securely
affixed to lid. Repair as needed.
7.
Verify oxygen regulator bleed port is installed to vent oxygen bleed gas through the
louvers in the inspiration access panel.
8.
Verify the check valves in the air intake manifold are securely installed in the air intake
manifold, and are not visibly worn or torn.
9.
Verify oxygen fitting is securely screwed into brass nut. (Brass nut will swivel, however.)
10. If you are regreasing the rack (every 15,000 hours and every time you install a new
piston/cylinder assembly), inspect the rack and pinion for wear, chips, or breakage.
11. Visually inspect ventilator exterior and interior for obvious problems, such as missing or
broken parts; loose assemblies; or disconnected wires, connectors, or tubing. Repair as
needed.
5.4 Preliminary ventilator setup
Set up the ventilator for the performance verification as follows:
1.
Install a complete Puritan Bennett™ (or equivalent) ventilator breathing circuit. Verify
bacteria filters are installed.
NOTE:
A humidification device is not required for the performance verification or ventilation.
2.
Connect ventilator to an oxygen source.
5.5 Preliminary pneumatic test equipment setup
To set up your pneumatic test equipment for use in the performance verification, verify all
equipment is calibrated and
sufficient warm-up time has elapsed prior to measurement.
5.6 Performance verification procedures
To ensure systematic performance verification and logical fault diagnosis, perform these tests
in the order given. If you need to repeat a test, however, the current control settings are
completely defined at the beginning of each individual check.
NOTE:
• If you are running the performance verification tests in order, you need only make
the ventilator settings shown in boldface.
• To locate the cause of a malfunction, note the boldfaced letter following the step
and refer to the corresponding index letter in Section 5.7 (“Troubleshooting”).