Battery Operation
7-6
Operator's Manual
2.
Switching to Backup Battery
3.
Running on Backup Battery
4.
Backup Battery Low
5.
Backup Battery Shutdown Imminent
In addition, the following are functional alarms for the battery system:
•
Power Pac Battery Temperature Alarm
•
Backup Battery Temperature Alarm
•
Backup Battery Failure Alarm
•
Backup Battery Low Charge Alarm
WARNING:
Immediately secure an external power source or insert a fully charged Power Pac battery pack
when a Backup Battery Shutdown Imminent alarm occurs.
Charge the Power Pac battery pack for a minimum of 3 hours for 100% recharge. If the battery
charge is not fully depleted, the charge time may be less.
7.8
Power Accessories
7.8.1
Power Pac Battery Pack (BAT3271A)
It is recommended to have extra Power Pac battery packs on hand at all times. The Power Pac is
hot swappable—one can be removed and another inserted without interrupting ventilation.
7.8.2
AC Power Supply (PWR3204P)
It is recommended to have an extra AC power supply available to charge the extra Power Pac
battery when not attached to the ventilator.
7.8.3
Country-Specific Power Cord
AC power supply cords with a region-appropriate outlet plug are available for order. Choose from
NA-North American style (PWR3207P), UK-British style (PWR3210P),or EU-European style
(PWR3211P).
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......