Pneumatics Panel Assembly and Components
Service Manual
4-19
4.16.2
Air and Oxygen Inspiratory Flow Sensors (GR-FTD2100P and GR-FTD2101P)
Figure 4-14.
Inspiratory Flow Sensors
To replace the flow sensors, refer to
and perform the following steps:
1.
Remove the pneumatics panel assembly by performing the procedure in section
, “
2.
Remove cable GR-CBL2116A from connector J1 on the flow sensor.
3.
Remove the flow sensors from the mixing block (BLK2107M) by pulling the flow sensor straight out of
the mixing block.
4.
Reinstall the flow sensors by reversing steps 2 and 3.
Note:
If either of the flow sensors are replaced, calibrations d1 or d2 must be performed. Refer to Chapter
for d1and
d2 calibration.
1
Mixing block (BLK2107M)
3
Air flow sensor (GR-FTD2100P)
2
Cable GR-CBL2116A connected to J1 on
flow sensors
4
Oxygen flow sensor (GR-FTD2101P)
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......