Front Panel Cover (CVR2104M)
Service Manual
4-13
4.12
Front Panel Cover (CVR2104M)
Figure 4-9.
Front Panel Cover (CVR2104M)
To remove and reinstall the front panel cover, refer to
and perform the following steps:
1.
Remove the display board by performing the procedure in section
2.
Using a 51/64 inch hex key, remove the set screw securing the adjustment knob to the encoder.
3.
Remove the adjustment knob.
4.
Remove the eight screws and washers securing the CVR2104M to the front panel bezel (BZL2108M).
5.
Remove the nut securing the flat ribbon ground cable to the side of the front panel cover (CVR2104M).
6.
Pull the front panel cover away from the front bezel and place in an ESD-safe container.
7.
Reinstall the front panel cover by reversing steps 2 through 6.
1
Screw (one of eight securing cover)
4
BZL2108M
2
Nut securing flat ribbon ground cable
5
Adjustment knob located on front panel
3
CVR2104M
6
Backlight inverter PCB
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......