14.
Press the
Manual Inflation
button to confirm the selection. The message “Calibration Complete” is
displayed.
15.
Reset the respiratory rate to 128.
16.
To calibrate the zero pressure offset of machine 2 pressure transducer (XD106), repeat steps 5 to 13, with
the following differences:
a.
Remove the green tubing from transducer XD106.
b.
Adjust the setting of
t Insp
to “2”.
c.
Calibrate the value of machine 2 to 0.00 ±0.05 cmH
2
O.
17.
To calibrate the zero pressure offset of machine 1 pressure transducer (XD105), repeat steps 5 to 13, with
the following differences:
a.
Remove the clear tubing from transducer XD105.
b.
Adjust the setting of
t Insp
to “4”.
c.
Calibrate the value of machine 1 to 0.00 ±0.05 cmH
2
O.
18.
To calibrate the gain of exhalation drive pressure (XD107), machine 2 (XD106), and machine 1 (XD105),
connect the low-pressure calibration test jig tool (TOL2108A) as follows:
a.
Connect the orange tube from TOL2108A to XD107 on GR-PCB2110A.
b.
Connect the green tube from TOL2108A to XD106 on GR-PCB2110A.
c.
Connect the clear tube from TOL2108A to XD105 on GR-PCB2110A.
d.
Connect one of the clear, large ID tubes to a calibrated pressure gauge.
e.
Connect the other clear, large ID tube to the pressure syringe (TOL1952P).
Note:
Reference
for tubing, transducer, and pressure gain values.
Table 5-3.
Gain Calibration
Gain setting
Transducer ID
LCD display
t Insp
Pressure
Tolerance
Designation
Tube color
XD107
Drive Pressure
1
110 cmH
2
O
±1.0
Bottom Port
Orange
XD106
Machine 2
3
110 cmH
2
O
±1.0
Bottom Port
Green
XD105
Machine 1
5
110 cmH
2
O
±1.0
Top Port
Clear
XD102
Air Gas Supply
7
25 psig
±0.1
Top Port
Yellow
XD103
O
2
Gas Supply
9
25 psig
±0.1
Top Port
Blue
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......