Preventive Maintenance
7-2
Operator’s Manual
7.3
Recycling and Disposal
When the monitoring system, battery, or accessories reach the end of useful life,
recycle or dispose of the equipment according to appropriate local and regional
regulations.
7.4
Battery Maintenance
WARNING:
Explosion hazard — Do not use the battery with other manufacturer's batteries,
different types or models of batteries such as dry batteries, nickel-metal hydride
batteries, or Lithium-ion batteries together.
WARNING:
Explosion hazard — Do not connect the battery reversed in positive (+) and negative
(-) terminals. Do not charge the battery with polarities reversed.
Caution:
Covidien strongly recommends recharging the battery when it has not been
recharged for six (6) or more months.
Caution:
Follow local government ordinances and recycling instructions regarding disposal or
recycling of device components, including batteries.
Caution:
Do not short-circuit the battery, as it may generate heat. To avoid short-circuiting, do
not let the battery come in contact with metal objects at any time, especially during
transport.
Caution:
Do not solder the battery directly. Heat applied during soldering may damage the
safety vent in the battery's positive cover.
Caution:
Do not deform the battery by applying pressure. Do not throw, hit, drop, fold or
impact the battery.
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......