Using Oximeter Softkey Menus
Operator’s Manual
51
Changing Adult and Neonate Default Settings
Table 7.
Adult and Neonate Default Settings
Option
Adult
Default Settings
Neonate
Default Settings
%SpO
2
Lower Alarm Limit
85%
%SpO
2
Upper Alarm Limit
100%
95%
Alarm Silence Duration
60 Seconds
Alarm Silence Duration OFF Setting
Disabled
Alarm Silence Reminder
Enabled
Alarm Volume
7 of 10
Display Contrast
Midrange
Display Format
GCF
Backlight Brightness
8 (Battery Power)
10 (AC Power)
Language
English
Pulse Beep Volume
4 of 10
Pulse Rate Lower Alarm Limit
40 BPM
90 BPM
Pulse Rate Upper Alarm Limit
170 BPM
190 BPM
Pulse Rate Delay
Off
Allow Pulse Rate Delay
Yes
Response Mode
Normal
Sensor Adjust Enable
Yes
SatSeconds
SPD enabled
100
SPD disabled
Off
Off
Allow SatSeconds
Yes
SPD
SPD enabled
1
Always Off (Disabled)
Allow SPD
Yes
No
Real-Time Trend Display
%SpO
2
Real-Time Trend Scale
15 Minutes
Trend Scale
8 Hours
Nurse Call Polarity
Normally Low
Data Port Baud Rate
9600
Data Port Protocol
ASCII
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......