Troubleshooting
104
Operator’s Manual
2.
Error messages
—Error messages provide information. The “Sensor Disconnected” error
message leaves any action to the discretion of the user. Advisory messages appear as
centered text at the bottom of the screen.
Figure 98.
Sensor Disconnected Message Screen
Table 16.
Common Prompts and Error Messages
Message
Time-out
in seconds
Exit on
Alarm
Exit on
Alarm
Silence
Displayed
Resolution
ADJUST CONTRAST
UP/DOWN
10
Yes
Yes
When pressing
and holding the
HELP/CONTRAST
key.
To adjust the contrast,
use the ADJUST UP to
increase the contrast or
ADJUST DOWN key to
decrease contrast.
CLOCK SETTING
LOST
None
No
No
If the N-600x
detects that the
real-time clock
has stopped
running and both
battery and AC
power are lost.
Power-cycle the oximeter.
If the error remains,
recharge or replace the
battery.
DATA IN SENSOR
5
No
Yes
When a sensor
containing data is
connected to the
oximeter.
Clears on time-out,
sensor disconnect, or
pressing the ALARM
SILENCE key, whichever
comes first.
DATA TYPE
SpO2+BPM
5
No
Yes
When a blank
event sensor is
connected to a
oximeter with
event data type
set to SpO
2
+BPM.
Clears on time-out,
sensor disconnect, or
pressing the ALARM
SILENCE key, whichever
comes first.
DATA TYPE: SpO2
5
No
Yes
When a blank
event sensor is
connected to a
oximeter with the
event data type
set to SpO
2
.
Clears on time-out,
sensor disconnect, or
pressing the ALARM
SILENCE key, whichever
comes first.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......