Manufacturer’s Declaration and Guidance
Service Manual
2-13
Enclosure Leakage Current
Condition
AC Line Polarity
Neutral
Line Cord
Power Line
Ground
IEC 60601-1
ANSI/AAMI 60601-1
Normal
Normal
Closed
Closed
100 μA
Single Fault
Open
Closed
500 μA
Closed
Open
Normal
Reversed
Closed
Closed
100 μA
Single Fault
Open
Closed
500 μA
Closed
Open
Table 2-10.
Earth and Enclosure Leakage Current Specifications (Continued)
Table 2-11.
Patient Applied and Patient Isolation Risk Current
Patient Applied Risk Current
Condition
AC Line Polarity
Neutral Line
Power Line
Ground Cable
IEC 60601-1
ANSI/AAMI 60601-1
Normal
Normal
Closed
Closed
100 μA
Single Fault
Open
Closed
500 μA
Closed
Open
Normal
Reversed
Closed
Closed
100 μA
Single Fault
Open
Closed
500 μA
Closed
Open
Patient Isolation Risk Current
Condition
AC Line Polarity
Neutral Line
Power Line
Ground Cable
IEC 60601-1
UL 60601-1
Single Fault
Normal
Closed
Closed
5000 μA
Reversed
Closed
Closed
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......