Repair
12-40
Service Manual
Figure 12-22.
LCD Assembly Removal
8.
Lift out the LCD assembly.
9.
Remove the display flex cable from the LCD assembly.
To replace the LCD Assembly
1.
Replace the new LCD assembly, orienting with the two smaller flex cables to the top.
2.
Thread the large, right-hand flex cable through the plastic LCD clamp plate, ensur-
ing the smaller top flex cable is free above the plastic LCD clamp plate and the
dog-leg flex cable threads through the plastic LCD clamp plate cut-away below
the battery.
3.
Tighten the three T-15 torx screws along the sides of the plate to 10 inch-pounds.
4.
Replace antennae PCBs, ensuring both cables remain unpinched and properly
routed. Reference