Power Components Replacement
Service Manual
12-31
12.6.4
Left Power Cable Assembly Replacement
To remove the left power cable assembly
1.
Disassemble the monitoring system chassis. Reference
2.
Move the front chassis off to the side.
3.
Identify the proper cables under the battery cradle. They are the cables to the
opposite side of the enclosure from the power entry module.
4.
Remove the five-pin Mini-mate black cable and connector from its J2 receptacle
from the power supply PCB.
5.
Remove the four-pin red and black (DC) cable and connector with ferrite bead
from its J2 receptacle on the battery interconnect PCB.
Figure 12-18.
Left Power Cable Assembly Replacement