Troubleshooting
11-2
Service Manual
appear in order of priority. To access suggestions for resolving the particular
message, press the MORE INFO button. If the monitoring system allows the
caregiver to dismiss the condition, pressing DISMISS ALARM clears the alarm,
but does not clear the condition until the caregiver takes appropriate action.
If the monitoring system detects a condition requiring caregiver intervention,
it displays a either a prompt or an error message with a recommended action.
Pressing ALARM SILENCE silences any audible tone for a period of up to two
(2) minutes. A countdown timer reflects any silence time remaining.
•
Prompts —
Prompts require a response. For example, the READY: ATTACH
SENSOR TO PATIENT AND MONITOR prompt reminds users to connect both an
interface cable and sensor to the monitoring system and to the patient.
Figure 11-1.
Ready Prompt
•
Alarms and Error condition messages —
When the monitoring system
detects an error condition, it displays the alarm message, suggests corrective
action, and sounds an alarm. It continues monitoring the patient. For example, the
SENSOR DISCONNECTED error message leaves any action to the discretion of the
user, but the PULSE RATE LOW error message requires immediate caregiver inter-
vention.
Figure 11-2.
Sensor Disconnected Message and Help Screen
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......