Performance Considerations
8-4
Service Manual
–
Venous pulsations
–
Dark skin pigment
–
Intravascular dyes, such as indocyanine green or methylene blue
–
Externally applied coloring agents (nail polish, dye, pigmented cream)
–
Defibrillation
8.4
Reducing EMI (Electromagnetic Interference)
WARNING:
EMI disruption can cause erratic readings, cessation of operation, or other
incorrect functioning.
WARNING:
The monitoring system is intended for use by healthcare professionals only.
It may cause radio interference or may disrupt the operation of nearby
equipment. Mitigation for such disruption may require re-orienting or
relocating the monitoring system or shielding the location.
Caution:
This device has been tested and found to comply with the limits for medical
devices related to IEC 60601-1-2: 2007. These limits are designed to provide
reasonable protection against harmful interference in a typical medical
installation.
Because of the proliferation of radio frequency transmitting equipment and other
sources of electrical noise in health care environments (for example, electrosurgical
units, cellular phones, mobile two-way radios, electrical appliances, and high-defi-
nition television), it is possible that high levels of such interference due to close
proximity or strength of a source might result in disruption of monitoring system
performance. Reference
The monitoring system is designed for use in environments in which electro-
magnetic interference might obscure the client’s pulse. During such interfer-
ence, measurements may seem inappropriate or the monitoring system may
not seem to operate correctly. EMI disruption can cause erratic readings, ces-
sation of operation, or other incorrect functioning. If this occurs, survey the site
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......