Data Output
Operator’s Manual Addendum
ADD-21
Table 1-3 describes the respiration rate data in the historical trend data output.
Refer to the
Operator’s Manual
for descriptions of the other data in the output.
Table 1-3.
Historical Trend Data Output Definitions (Respiration Rate Data Only)
Column
Description
RR Max
Maximum respiration rate value observed within the 1-sec. duration of the
record
RR Min
Minimum respiration rate value observed within the 1-sec. duration of the
record
RR LL
Lower alarm limit for respiration rate as configured by the user
RR UL
Upper alarm limit for respiration rate as configured by the user
Alarms
Encoded number indicating which alarms were active during the 1-sec.
duration of the record.
Note:
The alarm number is intended for interpretation by software such as
the Nellcor™ Analytics Tool (NAT). It is encoded as a mathematical OR of 32
unique numbers, each indicating a unique alarm. Numbers indicating respi-
ration rate alarms are listed below. Refer to the
Operator’s Manual
for other
alarms that can appear in the output. A value of 0 is reported if no alarms
were active.
0x00100000
RR Not Activated
0x04000000
Respiration Rate Low
0x08000000
Respiration Rate High