Nellcor™ Respiration Rate
ADD-16
Operator’s Manual Addendum
Figure 1-12.
ALARM LIMITS Menu Options
Note:
The initial alarm limits are the factory default settings or the institutional default
settings set by qualified service personnel. Any changes to these settings are
temporary. Limit changes remain in effect as long as the monitoring system retains
power, but return to the factory or institutional default settings at power off.
To temporarily change respiration rate alarm limits
1.
While in normal monitoring mode, press MENU.
2.
Press ALARM LIMITS.
3.
Press RESPIRATION RATE.
4.
For each limit, use the arrow keys or slide the bar up or down until reaching the
desired value. There must be a minimum of 5 breaths per minute between the
high and low limits.