Kangaroo

 

924

Enteral Feeding

 P

ump

Troubleshooting Guide

15 Hampshire Street

Mansfield, MA 

02048

1-800-962-9888
508-261-8000

www.covidien.com

 

COVIDIEN, COVIDIEN with Logo and ™ marked brands are 

trademarks of Covidien AG or its affiliate. 

©2008 Covidien AG or its affiliate. All rights reserved.

H-6201-xM-0108

SYMPTOM PROBABLE CAUSE

CORRECTIONS

Flow Error 
Alarm
(FLo Err)

Feeding container is empty.

Feeding tube or feeding set tubing 
is occluded.

Drip chamber is improperly placed 
in pump.

Drip chamber walls are coated 
with feeding formula.

Sensors in upper drip chamber 
guide are blocked or dirty. Drip 
chamber is too full.

Refill feeding container. If bag has been in use for 24 
hours, replace.

Locate point of occlusion and correct. Note: occlusion 
may occur upstream or downstream from pump.
Check to make sure roller clamp on feeding set tubing 
is open. Check the patency of feeding tube.

Check to make sure drip chamber is properly “locked” 
in position in drip chamber guide.

Check to make sure formula is not preventing sensors 
from proper operation. If formula cannot be removed 
from inside walls of chamber via chamber manipulation, 
replace set.

Check to make sure sensors are free of dried formula. 
Remove any deposits by using a cotton swab dampened 
with warm soapy water.

Low 
Battery 
Alarm

(Lo bAt)

Battery has been run down below 
point of maintaining accuracy of 
pump.

 Plug pump power cord into wall outlet.

Hold Error 
Alarm

(HLd Err)

Pump has been left in the hold 
mode for over 5 minutes.

Press        to silence alarm, then press        again to stop  
flashing, then press         again to restart pump.

No Set

(LoAd SEt)

Pump set has been improperly 
placed onto pump.

If "dots" appear with message; 
too much ambient light

Check to make sure only a Kangaroo

 924 Pump Set 

has been placed onto the pump.
Check to make sure that the Optix

 Connector on 

pump set has been properly positioned into  the Optix

 

lock on pump. Make sure there is no water or other 
contaminants on the Optix

 Connector or Optix

 lock 

window. 
Move Pump away from bright light.

System 
Error 
Alarm

(SYS Err)

Rotor is not turning while pump is 
running.

Rotor is turning but is not detected 
by rotor sensors.

Return pump for technical service.

Return pump for technical service.

No Audible 
Alarm

Battery is fully discharged.
Alarm circuitry failure.

Charge battery for a minimum of 12 hours.
Return pump for technical service.

START
HOLD

START
HOLD

START
HOLD