background image

 

Enteral Feeding

 P

ump

 

Directions For Use

  NORMAL OPERATION

  Plug the pump into AC wall outlet.
  BATTERY OPERATION

  Unplugging the pump from the AC line will place the pump on battery power. To  
  recharge the battery, plug the pump into an AC wall outlet. Pump will automati- 
  cally begin charging.
 

PROGRAMMING

-These general instructions are for inserting the Kangaroo

  

  924 Feeding Sets into the Kangaroo™ 924 Feeding Pump.
1.  Close roller clamp on pump set.
2.  Fill feeding bag or spike container prior to setting up pump.
3. 

In order to achieve proper accuracy, the fluid line in the feeding bag must be   

  28 inches above the top of the feeding pump when initiating the feeding cycle.

 

4.  Slowly open control clamp and fill entire line with fluid, leaving a few drops   
  of fluid in the drip chamber.     

 

  NOTE:  Avoid filling the drip chamber more than half full.
5.  Close roller clamp.
6.  Load set per arrow diagram on pump. 

Do not overstretch tubing.

7.  Press       and wait while pump performs a system check.
8.  Check VOL CLR light or press VOL then VOL CLR to clear volume.
9.  Set feed rate using one of the two following options:
  • Pressing the increase      or     decrease button, or
  • Turning the knob on the left side of the pump
 

Setting the Volume to be Delivered “VTBD” Rate (optional feature)

  • Press the VTBD button.
  • Set “VTBD” by using knob on side of pump, or by pressing the 
     increase      or      decrease button.  
• Only use the “VTBD” option if you want the pump to stop and alarm once a   
   set amount of formula is delivered.   
10 . Open roller clamp.
11. Press START/HOLD        . 

  TO CHANGE RATE:

1.  Press       .
2.  Set new rate by pressing the     or      keys on the front panel.
3.  Press       .

See pump operating manual for detailed instructions.

Kangaroo

™  

924

ON

t

t

t

t

t

t

START
HOLD

START
HOLD

START
HOLD

Reviews: