NORMAL OPERATION
Plug the pump into AC wall outlet.
BATTERY OPERATION
Unplugging the pump from the AC line will place the pump on battery power. To
recharge the battery, plug the pump into an AC wall outlet. Pump will automati-
cally begin charging.
PROGRAMMING
-These general instructions are for inserting the Kangaroo
™
924 Feeding Sets into the Kangaroo™ 924 Feeding Pump.
1. Close roller clamp on pump set.
2. Fill feeding bag or spike container prior to setting up pump.
3.
In order to achieve proper accuracy, the fluid line in the feeding bag must be
28 inches above the top of the feeding pump when initiating the feeding cycle.
4. Slowly open control clamp and fill entire line with fluid, leaving a few drops
of fluid in the drip chamber.
NOTE: Avoid filling the drip chamber more than half full.
5. Close roller clamp.
6. Load set per arrow diagram on pump.
Do not overstretch tubing.
7. Press and wait while pump performs a system check.
8. Check VOL CLR light or press VOL then VOL CLR to clear volume.
9. Set feed rate using one of the two following options:
• Pressing the increase or decrease button, or
• Turning the knob on the left side of the pump
Setting the Volume to be Delivered “VTBD” Rate (optional feature)
• Press the VTBD button.
• Set “VTBD” by using knob on side of pump, or by pressing the
increase or decrease button.
• Only use the “VTBD” option if you want the pump to stop and alarm once a
set amount of formula is delivered.
10 . Open roller clamp.
11. Press START/HOLD .
TO CHANGE RATE:
1. Press .
2. Set new rate by pressing the or keys on the front panel.
3. Press .
See pump operating manual for detailed instructions.
Kangaroo
™
924
ON
t
t
t
t
t
t
START
HOLD
START
HOLD
START
HOLD