STEP 2
Completely squeeze the handles together with one
hand while using the other hand to pull both toggle
levers back from the jaws until an audible click indicates
that the needle is properly loaded. Note that the red
indicator boxes are covered by the black button.
STEP 3
Gently press down on the tab located at the top of
the SULU. The jaws will release from the SULU. Pull the
SULU away from the jaws to unreel the suture from the
spool. When using longer lengths of suture, ensure that
the suture does not fall from the sterile field. Suture
SULUs are offered in 7" and 48"; other lengths can
be manually cut.
Test to ensure that the needle is fully seated by
passing the needle back and forth between the jaws
of the instrument. The instrument is now ready for
use. (If the needle does not pass smoothly, unload
the needle and reload.)
IndIcAtoR
boXes
| 6 |
LoAdIng the InstRUMent