background image

 

 

 

Adjusting the focusing tension  

The tension of the focusing knobs can be adjusted. 

You can set it from light to heavy according your 

own  preference.  Please  note  that  when  the 
specimen leaves the focus plane after focusing or 

the stage declines out of its own, then you need to 

adjust the tension 

To tighten the focusing knob (more heavy), rotate 

the tension adjustment ring counter-clockwise; to 

loosen it, please turn it in the clockwise direction 

Eyepieces  

Using a binocular (or trinocular) tube is less tiring 

for  the  eyes  than  a  monocular  tube.  In  order  to 

obtain  a  smooth  “compound”  image,  we 

recommend you to go through the below steps  

1. The interpupillary distance  

The correct interpupillary distance is reached when one round image is seen in the field of view (see image    

below). This distance can be set by either pulling the tubes towards each other or pulling them away from 

each other. This distance is different for each observer and thus should be set individually. When more users 

are working with the microscope it is recommended to remember your interpupillary distance for a quick set 
up during new microscopy sessions 

 

 

 

 

 

 

 

 

 

 

 

 

 

2. The correct eye point  

The eye point is the distance from the eyepiece to the user’s pupil. To obtain the correct eye point, move the 

eyes towards the eyepieces until a sharp image is reached at a full field of view 

3. Adjusting the diopter  

 

Set diopter adjustment ring to zero.  

 

Close the left eye and focus the right tube by adjusting the coarse- and fine adjustment knobs.  

 

Close the right eye and focus the left tube with the diopter adjustment ring 

 

This  procedure  should  be  followed  by  each  individual  user.  When  more  users  are  working  on  the  same 

microscope  it  is  recommended  to  remember  your  own  diopter  setting  for  a  quick  set  up  during  new 

microscopy sessions 

Abbe condenser  

Beneath the object stage an Abbe condenser N.A. 12.5 is mounted. The condenser can be adjusted in height 

by moving the rack and pinion knob beneath the mechanical stage. By adjusting the condenser you can focus 

Focusing tension 

Field of view before 

adjustment

 

 

 

Field of view after 

adjustment

 

 

 

Summary of Contents for 801854

Page 1: ...1 Microscope binocular 2 801854 User manual...

Page 2: ...rice performance ratio Please read this manual carefully before using this product to ensure correct and safe usage The contents of this manual are subject to change without notice The appearance of t...

Page 3: ...e focusing tension 9 Eyepieces 9 1 The interpupillary distance 9 2 The correct eye point 9 3 Adjusting the diopter 9 Abbe condenser 9 The field K hler diaphragm 10 Adjusting the Aperture Diaphragm 10...

Page 4: ...e lens regularly Cleaning optical components is a delicate matter Please read the cleaning instructions in this manual carefully Environment storage and use This product is a precision instrument and...

Page 5: ...move the microscope by its focusing knobs stage or head When needed move the microscope with two persons instead of one The S40x and S100x objectives are equipped with a spring mount to prevent damag...

Page 6: ...mechanical stage or clips G On Off switch not visible O Condenser with irisdiaphragm filterholder H Light intensity control P Coaxial stage controls I Eyepiece s Q Lamp housing J Dioptric adjustment...

Page 7: ...n not necessary but described for your convenience nonetheless Mounting the objectives 1 Rotate the coarse focusing knob to lower the stage to its lowest position 2 Install the objectives into the obj...

Page 8: ...t position 4 Close the iris diaphragm until it is just visible on the edge of the field of view The microscope is properly set for use with the 4x objective For each other magnification in bright fiel...

Page 9: ...be set individually When more users are working with the microscope it is recommended to remember your interpupillary distance for a quick set up during new microscopy sessions 2 The correct eye poin...

Page 10: ...ighest possible resolution depth of field and contrast 2 When contrast is low rotate the diaphragm adjustment ring to 70 80 of the N A of objective this will improve the contrast of the image The diap...

Page 11: ...2600 mA Maintenance and cleaning Always place the dust cover over your microscope after use Always keep the eyepiece and objectives mounted on the microscope to avoid dust entering the instrument Cle...

Page 12: ...ay blow In most cases replacing the fuse with the correct voltage will resolve the issue However should you encounter a blown fuse frequently please contact your distributor for further assistance Fus...

Reviews: