  2

No stimuli need to be activated.  To retrieve the data following the assessment period, 
you can choose to export the raw the data or create a simple analysis structure in this 
protocol.  Select the data element “N of DISCRETE EVENT(S) – TOTALS FOR RUN” 
and add one Bin for every Activity Monitor.  Choose one Activity Monitor (switch) for 
each Bin (see Figure 2). 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Once the heater of the H24-60 has reached a stable operating temperature - at least five 
(5) minutes - you can begin the Graphic State session to run the Activity Monitor 
assessment. 
 
 

EVALUATION / CONFIRMATION 
 

The Activity Monitors should produce data values between 

750

 and 

1,100

.  Values within 

this range indicate the Activity Monitor is operating correctly.  Should the data from an 
Activity Monitor fall marginally outside of this window we recommend that you run the 
calibration assessment again on the same monitor.  Should the data from the second 
calibration assessment fall outside of these values or if the data is well outside of these 

FIGRURE 2 