2
No stimuli need to be activated. To retrieve the data following the assessment period,
you can choose to export the raw the data or create a simple analysis structure in this
protocol. Select the data element “N of DISCRETE EVENT(S) – TOTALS FOR RUN”
and add one Bin for every Activity Monitor. Choose one Activity Monitor (switch) for
each Bin (see Figure 2).
Once the heater of the H24-60 has reached a stable operating temperature - at least five
(5) minutes - you can begin the Graphic State session to run the Activity Monitor
assessment.
EVALUATION / CONFIRMATION
The Activity Monitors should produce data values between
750
and
1,100
. Values within
this range indicate the Activity Monitor is operating correctly. Should the data from an
Activity Monitor fall marginally outside of this window we recommend that you run the
calibration assessment again on the same monitor. Should the data from the second
calibration assessment fall outside of these values or if the data is well outside of these
FIGRURE 2