5
During transfer, with patient in a sling attached to the lift, DO NOT roll caster base
over uneven surfaces that could cause the patient lift to tip over. Use steering handle
on the mast at all times to maneuver the patient lift.
1.3.5 Transferring the Patient
WARNING
When the patient is elevated a few inches off a stationary surface and before moving
the patient, check again to make sure that all hardware or straps are properly
connected. If any attachment is NOT properly in place, lower the patient back onto
the bed and rectify the issue.
Wheelchair wheels MUST be locked before lowering the patient into the wheelchair
for transport.
Make sure the wheelchair weight capacity can withstand the patient's weight prior
to loading the patient.
Be sure to check the sling attachments each time the sling is removed and replaced,
to ensure that it is properly attached before the patient is removed from the bed or
chair.
Mast pivot must be tight to ensure safe use of the patient lift. Bolt must be checked
at least every six months in conjunction with periodic maintenance.
1.3.6 Performing Maintenance
WARNING
ONLY qualified personnel should carry out the lift maintenance.
DO NOT attempt to open the pre-sealed actuator or obtain local service, for it will
VOID the warranty and might result in damage. Consult your dealer or manufacturer
for further information.
Inspect the casters and axle bolts every six months to check for tightness and wear.
Regular maintenance of patient lifts and accessories is necessary to ensure proper
operation. DO NOT overtighten mounting hardware. This will damage mounting
brackets.