i-MyRun.N Instruction Manual
Electrophoresis System
MyRunN/manual 04.10.28 10:48 AM ページ2
Page 1: ...SMO BIO CO LTD Toyo Ekimae Bldg 2 20 Toyo 2 Chome Koto ku Tokyo 135 0016 Japan Tel 81 3 5632 9605 Fax 81 3 5632 9614 i MyRun N Instruction Manual Electrophoresis System MyRunN manual 04 10 28 10 48 AM...
Page 2: ...IMR 201 Electrophorsis Tank for i MyRun N IMR 301 Gel Casting System for i MyRun N IMR 401 Support guide for Minigel L S IMR 501 Terms and Conditions of Sale All goods and services are sold subject t...
Page 3: ...sed in the operation of this instrument as it can develop sufficient voltage and current to produce a lethal shock To avoid any risk of injury the instrument should only be operated by properly traine...
Page 4: ...ld the power supply be connected to an electrophoresis unit that has a leakage path to earth the IMR 201 will detect this fault and the high voltage will be turned off 2 _Start current check To ensure...
Page 5: ...e for use on six lanes A1 A6 1 _Make the gels by using the 26 tooth comb on six lanes A1 A6 2 _Set a gel with gel casting plate on the electrophoresis tank 3 _Set the guide over the plate 4 _Load the...
Page 6: ...art of the gel unit with organic solvents such as ethanol or acetone 3 Power Supply Operation Check the Quick Start Operation Manual 1 _Insert the power supply connector into the electrophoresis tank...
Page 7: ..._The display shows the last value of the selected and keys 3 _Select and adjust the running time by the selecting the key followed by 4 _Select If you do not wish to use the timer 1 _After selecting t...
Page 8: ...centration is too high the temperature will increase rapidly If you want to withdraw the agarose gel from the support first cut the two small overhangs of agarose located at the end of the gel Note Re...