MicroQuark overview
MicroQuark is mainly made of:
•
MicroQuark unit
•
Flowmeter
•
Other accessories
Let us see an overview of the parts and their assembly.
MicroQuark unit
The unit is the core of the device, with the electronics and cables for its working.
The flowmeter
The turbine flowmeter is made of a handle (the reader) with a hole, in which it is placed the turbine.
The air passing through the helical conveyors, takes a spiral motion which causes the rotation of the turbine rotor. The rolling blade interrupts the
infrared light beamed by the two diodes of the reader. Every interruption represents 1/4 turn of the rotor, this allows to measure the number of
turn in the time.
For hygienic reasons, we strongly recommend the use of a bacterial filter, to be connected as in the picture by side.
Note:
While inserting the turbine, be sure to push the turbine up to touch the end of the reader.