![background image](http://html1.mh-extra.com/html/coselgi/melodia-cme-10-m/melodia-cme-10-m_user-instructions_2670344024.webp)
24
Problem
Possible cause
solution
The hearing
aids do not
respond with
a correspond-
ing change in
volume or
program to
the RC
a . The RC is used
beyond the trans-
mission range
b . Strong electro-
magnetic interfer-
ence in the vicinity
c . The RC battery is
dead or does not
work
d . The RC and the
hearing aids are
not matched
a . Move the RC closer
to the hearing aids
b . Move away from
known source of EM
interference
c . Change the RC bat-
tery
d . Make sure RC is
matched with hear-
ing aids (see User in-
structions for RC)