INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™
Page 10 of 48
1.4.
E
XTENDED
(D
UAL
) R
ANGE
O
PTION
1.4.1.
Operation
When the measuring range of the analyzer is larger than 1150BTU/SCF, an extended range
option is available which covers a Wobbe index of 2850.BTU/SCF. This is accomplished by
adding a second gas-mixing orifice and selection valves to make changeover possible.
The dilution ratios of each mixing orifice are chosen such that the measuring ranges overlap.
Via the software it is possible to create a 4/20mA current loop signal that covers the whole range.
It is necessary to establish a switch over point that must be calibrated.
For this reason the calibration system is expanded with an extra solenoid valve.
Summary of Contents for 9610
Page 27: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 26 of 48 ...
Page 32: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 31 of 48 3 3 2 Operation Menu ...
Page 34: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 33 of 48 3 3 4 Output Menu ...
Page 36: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 35 of 48 ...
Page 38: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 37 of 48 Display Menu ...
Page 39: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 38 of 48 3 3 7 Reset Alarms Menu ...
Page 46: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 45 of 48 5 INSTALLATION DRAWING ...
Page 48: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 46 of 48 ...
Page 50: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 48 of 48 ...
Page 51: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 48 of 48 ...