background image

4

5

2.

Prerinsing

Membranes fitted to Spin-X

®

UF concentrators contain trace amounts of glycerin and

sodium azide. Should these interfere with analysis they can be removed by rinsing fill
volume of buffer solution or deionized water through the concentrator. Decant filtrate
and concentrate before processing sample solution. If you do not want to use the pre-
rinsed device immediately, store it in the refrigerator with buffer or water covering
the membrane surface. Please do not allow the membrane to dry out.

3.

Sterilization of Polyethersulfone Membranes

Polyethersulfone membranes should not be autoclaved as high temperatures will
substantially increase membrane MWCO. To sterilize, use a 70% ethanol solution
or sterilizing gas mixture.

Technical Specifications

Table

1

. Technical Properties

Spin-X UF 6

Spin-X UF 20

Concentrator Capacity

Swing bucket rotor

6 mL

20 mL

Fixed angle rotor

6 mL

14 mL

Dimensions

Total length

122 mm

116 mm

Width

17 mm

30 mm

Active membrane area

2.5 cm

2

6.0 cm

2

Hold up volume of membrane

<10 µL

<20 µL

Dead stop volume*

30 µL

50 µL

Materials of Construction

Body

Polycarbonate

Polycarbonate

Filtrate vessel

Polycarbonate

Polycarbonate

Concentrator cap

Polypropylene

Polypropylene

Membrane

Polyethersulfone

Polyethersulfone

*Dead stop volume as designed in molding tool. This volume may vary depending on sample, sample

concentration, operation temperature and centrifuge rotor.

Table

2

. Maximum Recommended Spin Force (xg)

Swing Bucket

Fixed Angle

Spin-X

®

UF 6

5,000 to 50,000 MWCO PES

4,000 xg

10,000 xg

100,000 MWCO PES

4,000 xg

6,000 xg

Spin-X UF 20

5,000 to 50,000 MWCO PES

5,000 xg

8,000 xg

100,000 MWCO PES

3,000 xg

6,000 xg

Table

3

a. Performance Characteristics Spin-X UF 6

(Time to concentrate up to

30

x [min.] at

20

°C and solute recovery %)

Rotor

Swing Bucket

25° Fixed Angle

Start Volume

6 mL

6 mL

Min.

Rec.

Min.

Rec.

BSA 1.0 mg/mL (66,000 MW)

5,000 MWCO PES

20

98%

12

98%

10,000 MWCO PES

13

98%

10

98%

30,000 MWCO PES

12

98%

9

97%

IgG 0.25 mg/mL (160,000 MW)

30,000 MWCO PES

18

96%

15

95%

50,000 MWCO PES

17

96%

14

95%

100,000 MWCO PES

15

91%

12

91%

Table

3

b. Performance Characteristics Spin-X UF 20

(Time to concentrate up to

30

x [min.] at

20

°C and solute recovery %)

Rotor

Swing Bucket

25° Fixed Angle

Start Volume

20 mL

14 mL

Min.

Rec.

Min.

Rec.

BSA 1.0 mg/mL (66,000 MW)

5,000 MWCO PES

23

99%

29

99%

10,000 MWCO PES

16

98%

17

98%

30,000 MWCO PES

13

98%

15

98%

IgG 0.25 mg/mL (160,000 MW)

30,000 MWCO PES

27

97%

20

95%

50,000 MWCO PES

27

96%

22

95%

100,000 MWCO PES

25

91%

20

90%

Reviews: