4
5
2.
Prerinsing
Membranes fitted to Spin-X
®
UF concentrators contain trace amounts of glycerin and
sodium azide. Should these interfere with analysis they can be removed by rinsing fill
volume of buffer solution or deionized water through the concentrator. Decant filtrate
and concentrate before processing sample solution. If you do not want to use the pre-
rinsed device immediately, store it in the refrigerator with buffer or water covering
the membrane surface. Please do not allow the membrane to dry out.
3.
Sterilization of Polyethersulfone Membranes
Polyethersulfone membranes should not be autoclaved as high temperatures will
substantially increase membrane MWCO. To sterilize, use a 70% ethanol solution
or sterilizing gas mixture.
Technical Specifications
Table
1
. Technical Properties
Spin-X UF 6
Spin-X UF 20
Concentrator Capacity
Swing bucket rotor
6 mL
20 mL
Fixed angle rotor
6 mL
14 mL
Dimensions
Total length
122 mm
116 mm
Width
17 mm
30 mm
Active membrane area
2.5 cm
2
6.0 cm
2
Hold up volume of membrane
<10 µL
<20 µL
Dead stop volume*
30 µL
50 µL
Materials of Construction
Body
Polycarbonate
Polycarbonate
Filtrate vessel
Polycarbonate
Polycarbonate
Concentrator cap
Polypropylene
Polypropylene
Membrane
Polyethersulfone
Polyethersulfone
*Dead stop volume as designed in molding tool. This volume may vary depending on sample, sample
concentration, operation temperature and centrifuge rotor.
Table
2
. Maximum Recommended Spin Force (xg)
Swing Bucket
Fixed Angle
Spin-X
®
UF 6
5,000 to 50,000 MWCO PES
4,000 xg
10,000 xg
100,000 MWCO PES
4,000 xg
6,000 xg
Spin-X UF 20
5,000 to 50,000 MWCO PES
5,000 xg
8,000 xg
100,000 MWCO PES
3,000 xg
6,000 xg
Table
3
a. Performance Characteristics Spin-X UF 6
(Time to concentrate up to
30
x [min.] at
20
°C and solute recovery %)
Rotor
Swing Bucket
25° Fixed Angle
Start Volume
6 mL
6 mL
Min.
Rec.
Min.
Rec.
BSA 1.0 mg/mL (66,000 MW)
5,000 MWCO PES
20
98%
12
98%
10,000 MWCO PES
13
98%
10
98%
30,000 MWCO PES
12
98%
9
97%
IgG 0.25 mg/mL (160,000 MW)
30,000 MWCO PES
18
96%
15
95%
50,000 MWCO PES
17
96%
14
95%
100,000 MWCO PES
15
91%
12
91%
Table
3
b. Performance Characteristics Spin-X UF 20
(Time to concentrate up to
30
x [min.] at
20
°C and solute recovery %)
Rotor
Swing Bucket
25° Fixed Angle
Start Volume
20 mL
14 mL
Min.
Rec.
Min.
Rec.
BSA 1.0 mg/mL (66,000 MW)
5,000 MWCO PES
23
99%
29
99%
10,000 MWCO PES
16
98%
17
98%
30,000 MWCO PES
13
98%
15
98%
IgG 0.25 mg/mL (160,000 MW)
30,000 MWCO PES
27
97%
20
95%
50,000 MWCO PES
27
96%
22
95%
100,000 MWCO PES
25
91%
20
90%