| CORNING LSE 6L DIGITAL WATER BATH
6.0 Temperature Readjustment
This water bath has software for temperature readjustment. You must first measure the temperature in the middle of the chamber
or in the sample for a minimum of 2 hours, which is required to stabilize the temperature. You must measure the temperature with a
calibrated reference temperature measuring instrument with precision of 0.1°C or more. After 2 hours or more, read the temperature
on the thermometer and compare it with the temperature on the LED display. The difference between the thermometer reading and
the display is the value, which you will enter in the software.
Example 1:
Reference temperature is 37.9°C, temperature on display is 37°C. Difference is 37.9 - 37 = 0.9
Value is 0.9 (value you enter).
Example 2:
Reference temperature is 36.2°C, temperature on display is 37°C. Difference is 37 - 36.2 = -0.8
Value is -0.8 (value you enter).
Procedure for Temperature Readjustment:
Press the Up and Down arrow keys at the same time for more than 5 seconds until
Cor
(correction) appears on the display and Decimal
point blinks.
With the Up and Down arrow key, define the value (when Decimal point is blinking). Pay attention to the sign of value (plus or
minus). After should this be 5 times defining the temperature, the display will automatically show the temporary temperature.
7.0 Limited Warranty
Corning Incorporated (Corning) warrants that this product will be free from defects in material and workmanship for a period of
two (2) year from date of purchase. CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED, INCLUDING
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE. Corning’s sole obligation shall be to
repair or replace, at its option, any product or part thereof that proves defective in material or workmanship within the warranty
period, provided the purchaser notifies Corning of any such defect. Corning is not liable for any incidental or consequential damages,
commercial loss or any other damages from the use of this product.
This warranty is valid only if the product is used for its intended purpose and within the guidelines specified in the supplied instruction
manual. This warranty does not cover damage caused by accident, neglect, misuse, improper service, natural forces or other causes not
arising from defects in original material or workmanship. This warranty does not cover motor brushes, fuses, light bulbs, batteries or
damage to paint or finish. Claims for transit damage should be filed with the transportation carrier.
In the event this product fails within the specified period of time because of a defect in material or workmanship, contact Corning
Customer Service at: USA/Canada 1.800.492.1110, outside the U.S. +1.978.442.2200, visit
www.corning.com/lifesciences
, or contact
your local support office.
Corning’s Customer Service team will help arrange local service where available or coordinate a return authorization number and
shipping instructions. Products received without proper authorization will be returned. All items returned for service should be sent
postage prepaid in the original packaging or other suitable carton, padded to avoid damage. Corning will not be responsible for
damage incurred by improper packaging. Corning may elect for onsite service for larger equipment.
Some states do not allow limitation on the length of implied warranties or the exclusion or limitation of incidental or consequential
damages. This warranty gives you specific legal rights. You may have other rights which vary from state to state.
No individual may accept for, or on behalf of Corning, any other obligation of liability, or extend the period of this warranty.
For your reference, make a note of the serial and model number, date of purchase, and supplier here.
Serial No. __________________________________________
Date Purchased _______________________________________
Model No. _________________________________________ Supplier _____________________________________________