PI500
7-1
Coretronic Corporation
Chapter 7
Touch Function Test and Calibration
7-1
Touch Function Test
-
PC*2 : USB Port
Touch Driver
QC (p)99.exe
-
Touch Pen
-
Test Tooling (Acrylic)
2.
Notice
-
Don’t install another USB-Touch-Driver into the computer.
-
The Battery Voltage of Touch Pen must be higher than 1.2V.
-
You must prepare two PC, one is for test “QC (p)99.exe”, another one is for calibration.
-
If you change a Touch Pen in testing precess, you must re-plug in USB port.
-
The resolution must be set to 1024*768@75Hz.
1.
Equipment