Note!
The E00/E01 system relies on 200 to 240V AC to work, and touching the line voltage can
cause fatal injuries!
If the fuse fails and the entire system is not working, unplug the power cord from the E00/E01
system and replace it with a new one.
The fuse is model AC250V-3A and the power supply connection and fuse are located on the right
side of the rear panel of the E00/E01 chassis.
If the system is still not working properly after replacing the new fuse, please contact your sales
engineer or our after-sales service department immediately. The steps to remove the fuse are as
follows:
1) Turn off the E00/E01 system and unplug the power cord.
2) After waiting for one minute, make sure any residual voltage has dissipated.
3) Open the holder that covers the fuse and pull out the fuse holder.
4) Replace the fuse of the same type and place it on the fuse holder.
5) Put the fuse holder back in place.
6.3 Transportation and storage
This product is packed in carton. Transportation must be carried out under product packaging
conditions, and direct rain and snow, direct contact with corrosive gases and strong vibrations
should be avoided during transportation.
The instrument can be transported under various conditions of normal transportation, and should
avoid damp, load, collision, extrusion, irregular placement and other adverse conditions during
transportation.
If the instrument is not used for a long time, the instrument should be packaged and stored.
The instrument should be stored in a non-corrosive atmosphere and in a well ventilated, clean
room.
In the process of transportation, storage and use, attention should be paid to fire prevention,
shockproof, waterproof and moisture proof.