16
17
Troubleshooting
Problem
Cause
Solution
The CONTROL unit does not
seem to be charging
Charging cable not plugged in
fully.
Ensure the USB cable is firmly in
the CONTROL unit and other end
is in the AC Adapter or Computer
USB Port. Check if there is a
problem with the AC adapter by
plugging the USB Cable into a
computer port.
Skin is red in treatment area
Overtreatment or adverse
reaction
Stop and treat another area.
Contact your physician if problem
persists.
Stimulation is weak or non-
existent
Pad(s) not adhering well or
battery is low
Clean skin
Replace electrode pad(s)
Charge battery
Specifications
Size:
5.6x4.4x 1.4cm
Weight:
20g
Power supply:
DC 3.7V Lithium rechargeable battery
Channels:
Single channel
Waveform:
Asymmetric biphasic pulse
Intensity:
0-80 mA @ 500 Ω
Pulse Width:
100μS
Frequency:
1-52 Hz
Timer:
Continuous for 3 hours duration period; unit automatically it shuts
off at completion
Intensity:
20 levels/adjustable
Mode:
2 modes; automatically adjusts
Operating Condition:
5
o
C to 40
o
C (41
o
F to 104
o
F); RH 30% - 75%; AP 700 to 1060 Hpa
Storage condition:
-10
o
C to 55
o
C (14
o
F to 131
o
F); RH 10% - 90%; AP 700 to 1060 Hpa
Summary of Contents for ELT-2700
Page 2: ...2 User Manual English...
Page 19: ...19...
Page 20: ...20 Manuel de l utilisateur Fran ais...
Page 37: ...37...
Page 38: ...38 Manual del usuario Espa ol...
Page 55: ...55...