CALC:TRAN:TIME:KBES
CALCulate<Ch>[:SELected]:TRANsform:TIME:KBESsel
<numeric>
CALCulate<Ch>[:SELected]:TRANsform:TIME:KBESsel?
Description
Sets
or
reads
out
the
β
parameter,
which
controls
the
Kaiser–Bessel
window
shape,
when
performing
time
domain
transformation
(command/query).
Target
The
active
trace
of
channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
<
numeric
>
β
parameter
from
0
to
13
Out
of
Range
Sets
the
value
of
the
limit,
which
is
closer
to
the
specified
value.
Query
Response
<numeric>
Preset
Value
6
Equivalent
Softkeys
Analysis
>
Time
Domain
>
Window
>
Kaiser
Beta
CALC:TRAN:TIME:LPFR
CALCulate<Ch>[:SELected]:TRANsform:TIME:LPFRequency
Description
Ch
anges
the
frequency
range
to
match
with
the
lowpass
type
of
the
time
domain
transformation
function
(no
query).
Target
The
active
trace
of
channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Equivalent
Softkeys
Analysis
>
Time
Domain
>
Set
Frequency
Low
Pass