Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
198
SENS:ROSC:SOUR
SENSe<Ch>:ROSCillator:SOURce
{INTernal|EXTernal}
SENSe<Ch>:ROSCillator:SOURce?
Description
Sets
or
reads
out
the
internal
or
external
source
of
the
reference
frequency
of
10
MHz
(command/query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
"INTernal"
:
Internal
source
of
the
reference
frequency
"EXTernal"
:
External
source
of
the
reference
frequency
Out
of
Range
Error
occurs.
The
command
is
ignored.
Error
code:
224.
Query
Response
{INT|EXT}
Preset
Value
INT
Equivalent
Softkeys
System
>
Misc
Setup
>
Ref
Source