Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
161
SENS:CORR:COLL:CKIT:STAN:Z0
SENSe<Ch>:CORRection:COLLect:CKIT:STAN<Std>:Z0
<impedance>
SENSe<Ch>:CORRection:COLLect:CKIT:STAN<Std>:Z0?
Description
Sets
or
reads
out
the
offset
Z0
value
for
the
calibration
standard
(command/query).
Target
Standard
<Std>
of
the
calibration
kit
specified
for
channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
<Std>={[1]|2|...N},
where
N
–
the
number
of
the
standards
in
the
calibration
kit
Parameter
<impedance>
the
offset
Z0
value
from
–1E18
to
1E18
Unit
Ω
(Ohm)
Out
of
Range
Sets
the
value
of
the
limit,
which
is
closer
to
the
specified
value.
Query
Response
<numeric>
Preset
Value
50
or
75,
depending
on
the
selected
calibration
kit
Equivalent
Softkeys
Calibration
>
Edit
Cal
Kit
>
Offset
Z0
SENS:CORR:COLL:CLE
SENSe<Ch>:CORRection:COLLect:CLEar
Description
Clears
the
measurement
values
of
the
calibration
standards
(no
query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Equivalent
Softkeys
Calibration
>
Calibrate
>
{
Response
(Open)
|
Response
(Short)
|
Response
(Thru)
|
One
Path
2–Port
Cal
|
Full
1–Port
Cal
}
>
Cancel